10 cc Ion Chamber
Model: 7047
Application
The 10 cc Ionization Chamber is designed as a high activity, bakeable radiation detector. With the small footprint and VCR8 connections the chamber is easily incorporated into any process. The specialized ceramic feed through allows a bias of up to 100V and an anode signal as low as 1 fA (corresponds to 100 µCi/m³ of tritium). When combined with Tyne’s preamplifier and controller package or signal processor this makes an ideal tool for measuring tritium gas in any process.
Note: For lower tritium concentrations, please see Tyne's 1000cc ion chamber.
Features
-
10cc ion chamber suitable for mCi/m³ to kCi/m³ of Tritium
-
Virtual Wall for minimum tritium surface contamination
-
Can be used in explosive environment with purge equipment
-
Precise, tested and proven 10cc geometry.
-
Can bias anode up to 100V from ground
-
Inlet precipitator
-
Bakable to cleanse contaminated surfaces from ionization chamber
-
Use in a process line or for room air measurements
-
Full vacuum to 150 psig
-
Flows up to 1 l/s
Description
The 10 cc Ion Chamber is designed to be an in-line radiation in a process gas detector with the most typical application being Tritium. The smaller volume from the 1000cc detector and the virtual wall technology makes this the ideal choice for very high activity measurements. The inlet and outlet VCR 8 connectors allow the chamber to be easily mounted into an existing process line. The internal anode is connected to a floating BNC which can be biased up to 100 V. The high resistance ceramic of the feed through allows the ionization chamber to measure ionization currents as low as 1 fA (1 µCi/m³ of Tritium). The bias voltage is also connected to an inlet precipitator to remove any ionized particles carried into the chamber. The maximum inlet flow rate of 1 L/s ensures that all particles are neutralized before entering the chamber. The ionization chamber is an approved pressure vessel that can handle up to 150 psig, is bakeable up to 350°C and is 1x10E-9 Helium leak tight.
The ionization chamber is typically connected to the Tyne Engineering preamplifier or signal processor. The preamplifier is a multi-ranged amplifier that connects to a remote controller which gives the user the maximum flexibility for measuring the amount of tritium, auto zeroing and alarm functions. The signal processor is a low cost device that provides the bias and converts the 1 fA signal to an easily handled 0 – 10 V signal that can be integrated into any existing data acquisition system.
Specifications
Volume | 10 cc |
Measurement Type | Measure ions with a biased Anode |
Detection volume Type | Virtual Wall – low activity hang up |
Maximum Bias | 100 VDC |
Minimum Measurable Current | 1 fA (equivalent to 0.1 mCi/m³ of Tritium) |
External Gamma source Signal | 5 fA per 1 mR/h |
Carry over Protection | Virtual wall precipitator |
Flow rate | 1 L/s |
Pressure | Vacuum to 150 psig |
Temperature | 350 °C |
Leak Tightness | 1 x 10E-8 cc/sec |
Relative Humidity | 95 % |
Interior | Electro polished |
Wetted Parts | 316L Stainless Steel, High Density Ceramic |
QA | ISO 9001:2000 |